Constant temperature and humidity box operation guide

Constant temperature and humidity test chamber operation guide:

1. The purpose of the request is to make the instrument operate normally and ensure the reliability of the results.

2. Requirements for the scope of laboratories

3. The responsible laboratory technician shall strictly abide by the operating procedures, and the QC supervisor shall be responsible for supervising the implementation of this regulation.

4. Definition no

5. Safety Precautions This instrument requires an external water pipe, and care should be taken to prevent water leakage during use.

6. Procedure

6.1. In order to extend the instructions of the instrument and achieve good control results, the inlet water of the constant temperature and humidity instrument must meet the following conditions:
• The water supply cannot be interrupted during the operation of the instrument.
· The water supply pressure must be between 1 and 10 bar.
• The water supply conductivity must be between 200 μs/cm and 500 μs/cm.
· The water supply temperature should be below 40 °C.

6.2. Temperature control
6.2.1 Turn on the temperature control switch and press [ENTER] to the temperature control state.
6.2.2 Press the [Pgm] button, use the [s][t] buttons to set the temperature, and press [ENTER] to confirm.
6.2.3 Press the [EXIT] button to return to the main menu.

6.3. Humidity control
6.3.1 Turn on the humidity control switch and press [ENTER] to the humidity control state.
6.3.2 Press the [Pgm] button, use the [s][t] buttons to set the humidity, and press [ENTER] to confirm.
6.3.3 Press the [EXIT] button to return to the main menu.

6.4. When carrying out the accelerated experiment, first read the quality standard of the product carefully, select the temperature and humidity conditions of the accelerated test, and then set the required temperature and humidity values ​​according to the methods of 6.2 and 6.3.

6.5. During the accelerated experiment, the temperature and humidity of the constant temperature and humidity meter should be recorded daily. If any abnormal conditions are found, report it to the laboratory supervisor for maintenance.